NT2C Vital Signs Monitors
Download 1:Product material one
Product Name:NT2C Vital Signs Monitors
Product Feature |
|
----------------------------------------------------------- | |
◆Bright, Easy-to-Read Large LED Display | |
◆Durable, Compact and Lightweight | |
◆Long-life Internal Rechargeable Battery for Uninterrupted Monitoring | |
◆Audible/Visual Three Level Alarms with Alarm Volume Control | |
◆72-hour Data Memory Allowing On-screen Viewing of SpO2 and Pulse Rate Trends | |
◆Suitable for Adult, Pediatric and Neonate | |
----------------------------------------------------------- | |
Product Specification | |
◆Performance Measurement Range: | ◆Electrical Instrument: |
SpO2: 0%~100% | Power Requirements: 100~240VAC, 50/60Hz, 40VA |
Pulse Rate: 30bpm~250bpm | Fuse Rating: 1.0A |
◆SpO2 Accuracy: | ◆Battery: |
Adults, Pediatrics: 70% ~ 100% ±2% | Internal Battery: 12V |
0%-69% unspecified | Battery Capacity: ≥ 24 hours |
Neonates: 70% ~ 100% ±3% | |
0%-69% unspecified | ◆Environment: |
◆Pulse Rate Accuracy: | Operating: |
Adults, Pediatrics, Neonates: ±3% | Temperature: 5OC~40OC (41OF ~104OF) |
Humidity: 10%~95% | |
◆Non-Invasive Blood Pressure: | Transport/Storage: |
Measurement type: Adult, Pediatric, Neonatal | Temperature: -20OC~70OC (-4OF ~158OF) |
Measurement range: | Humidity: 10%~95% |
Adult: Systolic 40 ~ 270mmHg | |
Diastolic 20 ~ 200mmHg | ◆Physical Characteristics: |
Mean 30 ~ 220mmHg | Dimensions: 264mm(W) x 95mm(H) x 172mm(D) |
Pediatric: Systolic 40 ~ 180mmHg | 10.39in(W) x 3.74in(H) x 6.77in(D) |
Diastolic 15 ~ 160mmHg | Net Weight: 2.25kg/5 Ib |
Mean 20 ~ 170mmHg | |
Neonatal: Systolic 30 ~ 140mmHg | ◆Compliance: |
Diastolic 15 ~ 100mmHg | 1. SpO2: ISO 9919: 2005(E) |
Mean 20 ~ 115mmHg | 2. Safety Standards: |
Accuracy: Meets ANSI/AAMI SP10-1992 | EN 60601-1: 1990+A1: 1993+A2: 1995+A13: 1996 |
Resolution: 1mmHg | 3. Alarm: IEC60601-1-8: 2005 |
Protection: over pressure | 4. EMC: EN 60601-1-2: 2007, Group 1 Class A |
5. Environment : WEEE (2002/96/EC) | |
◆Alarm | |
Audible/Visual Alarms | ◆Output: |
Adjustable Alarm Limits | RS-232 Serial Port for Digital Output |
Nurse Call Function | |
◆Temperature: | |
Measurement range: 25 ~ 45 ℃ | |
Accuracy: ± 0.2 ℃ |